12-2 Care & Maintenance
NO.
Maintain content
Frequency
Method
3.
Clean trackball
Once a month
Same as the above
4.
Clean control panel and minor panel
Once a month
Same as the above
5.
Clean probes (the head)
Every time after
using
Same as the above
6.
Clean probe cable and the surface of
connector
Once a month
Same as the above
7.
Clean holders (including probe holder
and gel holder)
Once a month
Same as the above
8.
Clean cover
Once a month
Same as the above
9.
Clean peripherals
Once a month
Referring to12.2.2
10.
Check surface of probes
Once a day
Referring to12.3.1
11.
Check power cable , plug and circuit
breaker
Once a month
Referring to12.3.1
12.
Check battery
Once a year
Referring to12.3.1
13.
Check function of peripherals and
options
Once a year
Referring to12.3.3
14.
Mechanical safety inspection
Once a year
Referring to12.3.4
15.
Electrical safety inspection
Once a year
Referring to appendix A

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish

Related product manuals