Description of Self-diagnosis Test Items C-21
If prompt “ECG Module is not on the site.”states that ECG is not in place or in-place detection
circuit is failed; if prompt “ECG Module information read test FAIL.”there may be communication
malfunction between ECG board and DSP board.
Suggestion when test failed
Check if ECG board, back-end mother board, DSP board and digital board. First option is to change
ECG board.
C.1.45 Z1502 ECG Module Self Test
Test Content
The super stratum test itemZ1401 ECG Module Information Read Test.
Check if ECG board is in place, send order to let the ECG do self test and waiting the feedback
order.
Analysis when test failed
If prompt “ ECG Module is not on the site.”states that ECG is not in place or in-place detection
circuit is failed; if prompt “ECG Module self test FAIL;”there may be malfunction on ECG board.
Suggestion when test failed
Replacing ECG board is recommended.