6-30 Cardiology
Operating Procedures
1. Select [RVSP] in the measurement menu.
2. Measure TR Vmax in Doppler mode.
The TR PGmax is calculated.
3. Select [RAP] in the [RVSP] sub-menu, and select (or enter) the pressure in the dialog box
popped up. As shown in figure below:
Range of input values is [0, 50.0mmHg].
4. Click [OK] after selecting (or inputting) the pressure, the RAP is obtained.
RVSP is calculated.
6.4.3.9 PAEDP
PAEDP measures the pulmonary artery end diastolic pressure.
Study Items
Tools Descriptions Operations
PR Ved
Pulmonary Valve Regurgitation
Velocity at end-Diastole
D Vel in Doppler General Measurements
RAP Right Atrium Pressure See RAP measurement in “RVSP”
Study Results
Except for values in upper table, the following results can be obtained in this study:
Tools Descriptions Formulae
PR PGed
Pulmonary Valve Regurgitation
Pressure Gradient at end-
Diastole
/
PAEDP
Pulmonary Pressure at end-
Diastole
2
))/( (4)()PAEDP(mmHg smVedPRmmHgRAP ×+=
Operating Procedures
See table above for methods and formulae of the measurement items.
6.4.3.10 RV TEI (RVIMP)
Measurement of RVIMP (Right Ventricular Index of Myocardial Performance) is similar to that of