9-6 Urology
PSAD (ng/ml
2
) = Serum PSA (ng/ml) / Prostate Vol (ml)
Here, PPSA Coefficient and Serum PSA are input in [Patient Info] -> [URO] dialog box. The default
value of PPSA Coefficient is 0.12.
1. Select the [Prostate] in the measurement menu.
2. Use the Distance method in 2D General Measurements to measure Prostate L, Prostate H
and Prostate W.
The system calculates Prostate Vol and PPSA.
The PSAD displays in the report if the PSA value is input.
Seminal Vesicle
Function: measures Seminal L, Seminal H and Seminal W.
NOTE: Need to be measured in left and right side respectively.
1. Select the [Seminal Vesicle] in the measurement menu.
2. Use the method of Distance measurement of 2D General Measurements to measure Seminal
L, Seminal H and Seminal W.
Testis
Function: measures Testis L, Tes ti s H a nd Tes tis W, calculates Testis Vol.
NOTE: Need to be measured in left and right side respectively.
1. Select the [Testis] in the measurement menu.
2. Use the method of Distance measurement of 2D General Measurements to measure Testis L,
Testis H and Testis W. The Testis Vol is calculated automatically.
Bladder
Function: measures Pre-BL L, Pre-BL H, Pre-BL W, Post-BL L, Post-BL H and Post-BL W,
calculates Pre-BL Vol, Post-BL Vol and Mictur.Vol.
1. Select the [Bladder] in the measurement menu.
2. Use the Distance method in 2D General Measurements to measure Pre-BL L, Pre-BL H and
Pre-BL W. The Pre-BL Vol is calculated automatically.
3. Use the Distance method in 2D General Measurements to measure Post-BL L, Post-BL H and
Post-BL W. The Post-BL Vol and Mictur.Vol is calculated automatically.
Mass
Function: measures 3 distances in the mass.
Take Mass1 as an example, the procedures are as follows:
1. Select [Mass1] in the measurement menu.
Rotate the knob under [Position] on the touch screen to record the mass position.
2. Use the Distance tool in the 2D General Measurement to measure Mass1 d1, Mass1 d2 and
Mass1 d3.
Epididymis
Function: measures Epididymis L, Epididymis H and Epididymis W.
NOTE: Need to be measured in left and right side respectively.
1. Select the [Epididymis] in the measurement menu.
2. Use the method of Distance measurement of 2D General Measurements to measure
Epididymis L, Epididymis H and Epididymis W.

Table of Contents

Other manuals for Mindray DC-8 EXP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-8 EXP and is the answer not in the manual?

Mindray DC-8 EXP Specifications

General IconGeneral
BrandMindray
ModelDC-8 EXP
CategoryMedical Equipment
LanguageEnglish

Related product manuals