EasyManua.ls Logo

Mindray DP-10 - Connecting Equipotential Terminal; Connecting Transducer

Mindray DP-10
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Connection
2-2
WARNING:
When the system is connected with other equipments, be
sure to connect their equipotential terminals before inserting
any of their power plugs into a receptacle, and also, be sure
to remove the power plugs from receptacles before
disconnecting the equipotential terminals.
2.3 Connecting Transducer
WARNING:
Use only transducers specified in this manual. Otherwise, the
system and transducer may be damaged or an accident such
as a fire may occur.
Ensure the transducer and its connector and cable are intact
before connecting it to the system. Using defective
transducers may cause electric shock.
NOTE:
Before connecting or disconnecting a transducer, be sure to turn off the
system or stop emitting (freeze the image). Otherwise, the system or
transducer may be damaged.
To connect a transducer
1 Turn off the system or freeze the image.
2 Put the lock handle to the position shown in the figure below.
3 Plug the transducer connector into the transducer socket on the system.
4 Turn the lock handle 90° clockwise.
To disconnect a transducer

Table of Contents

Related product manuals