Transducers and Biopsy
12-19
Plastic/needle detachable needle-guided bracket:
1
2
3
4
5
No. Name Description
1 Guiding block
Used for installing biopsy needle; there are five
specifications of guiding blocks for different biopsy needles
2 Angle block
Used for determining the angle of the biopsy; there are
three specifications of blocks of angle
3
Support of
needle-guided
bracket
Used for installing the needle-guided bracket on the
transducer
4
Specification of
guiding block (13G)
Matched with the corresponding biopsy needle (13G)
5
Groove and tab of
the needle-guided
bracket
Respectively matched with the tab and groove of the
transducer
6 Needle guide hole Used for installing the biopsy needle
7
Specification of angle
block (60°)
Corresponding to the size of the biopsy angle (60°)
12.2.3 Inspection of the Needle-guided Bracket
Be sure to perform inspections before and after use of the needle-guided bracket. If an
abnormality is found on the needle-guided bracket, immediately stop using it and contact
MINDRAY Customer Service Department or sales representative.
1. Sterilize the needle-guided bracket before and after use.
2. Confirm that the needle-guided bracket is free of damage, deformation, stripping,
malfunction, loose, or missing parts.
3. Confirm that the needle-guided bracket is securely mounted in the correct position.
12.2.4 Installing the Needle-guided Bracket
NGB-001 metal/needle detachable needle-guided bracket
1) Put on the transducer cover.
2) Select a proper needle-guided bracket, and match the groove with the tab of the
transducer. Mount the bracket onto the transducer.