3-1
3 Startup/Shutdown
This chapter introduces procedures of startup and shutdown of the system.
3.1 Startup
1 Before the startup, complete the following checks:
The ambient temperature, humidity and atmospheric pressure meet requirements
listed in the appendix.
The system is placed in a clean location where there is no direct sunlight, no sudden
changes in temperature, no condensation, few dust, no vibration, and far from
heat generators. Ensure at least 20cm of clearance at the back and sides of the
system.
The system and peripherals have no deformation, damage or dirt. If necessary, clean
the system (see the “System Maintenance” chapter for details).
There are no obstacles in the area where the system can be moved and near the
radiator holes of the system.
All power cables and connection cables are intact.
The transducer(s) has(have) been cleaned and disinfected (see transducers’ manuals
for details).
The transducer(s), connector(s) and cable(s) are intact.
The transducer(s) is(are) connected to the system.
Peripherals to be used are connected to the system.
All connections are secure.
2 Ensure the power plug of the system is connected to an outlet meeting requirements.
For peripherals to be used, ensure their power plugs are connected to receptacles
meeting requirements.
3 Turn on the system. For peripherals to be used, turn them on.