Chapter 1 About the Product
3
SEV (Sevoflurane)
DES (desflurane)
Airway Respiration Rate (rpm: Respiration Per Minute): AwRR
Minimum Alveolar Concentration (MAC)
4 AG waveforms (CO
2
, N
2
O, O
2
, AA)
The PM-7000 provides the functions of audio/visual alarm, trend graphic storage and output,
NIBP measurement, alarm event identification, large font screen, defibrillator synchronization,
oxyCRG recall, drug calculation, etc.
1.3 Environment
1.3.1 Temperature
Work mode 0 ~40
MINDRAY CO
2
module +5 ~ +35
Welch Allyn mainstream CO
2
module +10 ~ +35
Microstream CO
2
module +5 ~ +35
Artema AION AG module +10 ~ +35
Transportation & Storage -20 ~ 60
1.3.2 Humidity
Work mode 15% 95 % (non-condensing)
Transportation & Storage 10% 95 % (non-condensing)
Atmospheric pressure 70.0kPa 106.0kPa
1.3.3 Electrical specification
AC power supply: 100 to 240 V AC, 50/60 Hz, Maximum input power: 140VA
DC power supply: 12 V (nominal), 10 to 16 V, Power:
80W
2.3 Ah 12V lead-acid rechargeable battery
Working time of fully-charged batteries in normal status: 75 minutes (1 battery).
From the first low-battery alarm, the batteries can supply power to the patient monitor for 5
~ 15 minutes.
Maximum charging time: about 8h
4.4Ah 11.1V lithium battery
Working time of fully-charged batteries in normal status: 150 minutes (1 battery).
From the first low-battery alarm, the batteries can supply power to the patient monitor for 5
~ 15 minutes.
Maximum charging time: about 8h