Chapter 3 Product Specifications
35
3.10 SpO
2
Specifications
1.1.1. Mindray SpO
2
Module
SpO
2
Measurement range 0 to 100%
Resolution 1%
Precision
70 to 100%:
70 to 100%:
70 to 100%:
0% to 69%:
±2 % (adult/pediatric, non-motion conditions)
±3 % (neonate, non-motion conditions)
±3 % (in motion conditions)
Undefined.
Refreshing rate 1s
PR
Measurement range 20 to 254bpm
Resolution 1bpm
Precision
±3 bpm (non-motion conditions)
±5 bpm (in motion conditions)
Refreshing rate 1s
1.1.2. Masimo SpO
2
Specifications
SpO
2
Measurement range 1 to 100%
Resolution 1%
Precision
70 to 100%:
70 to 100%:
70 to 100%:
0% to 69%:
±2% (adult/pediatric, non-motion conditions)
±3% (neonate, non-motion conditions)
±3% (in motion conditions)
Undefined.
Refreshing rate 1s
PR

Table of Contents

Related product manuals