Chapter 3 Product Specifications
39
3.13 IBP Specifications
Number of channels 2
Pre s su r e la b els ART, PA, C VP, R A P, LA P, IC P, P1 a nd P 2
ART 0 to 300 mmHg
PA – 6 to 120 mmHg
CVP/RAP/LAP/ICP – 10 to 40 mmHg
Measurement range
P1/P2 – 50 to 300 mmHg
Resolution 1 mmHg
Precision ±2% or ±1mmHg, whichever is greater
Update period 1s
Pressure transducer
Sensitivity 5 uV/V/mmHg
Impedance range 300 to 3000