EasyManua.ls Logo

Mindray PM-7000 - CO Specifications; Mindray CO

Mindray PM-7000
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Product Specifications
41
3.15 CO
2
Specifications
Measurement technique Infrared absorption technique
Measurement mode Sidestream, microstream or mainstream (optional)
Displayed parameter EtCO2, FiCO2, Respiration Rate
CO2 function Meet the requirements of EN864 and ISO9918.
1.1.4. Mindray CO
2
Specifications
CO2 measurement range 0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2mmHg
±5%
±10%
Deflation rate 100, 150ml/min
Precision of deflation rate 15%
Start-up time of CO2
module
< 1min.
The module enters the warming up status after the startup. Ten
minutes later, it enters the ready-to-measure status.
AwRR measurement range 0 to 120 BrPM
Precision
0 to 70 BrPM:
> 70 BrPM:
±2 BrPM
±5 BrPM
Response time < 240 ms (10 to 90%)
Delay time
< 2s (Length of sampling tube: 7 feet; inner diameter: 0.055
inches; sampling flow: 150ml/min)
Apnea alarm delay AwRR: 10 to 40 s

Table of Contents