EasyManua.ls Logo

Mindray PM-7000 - Ag Calibrate; AG Check

Mindray PM-7000
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Test and Material List
64
5.5 AG CALIBRATE
5.5.1 AG Check1Using T-piece to connect the watertrap and Agent steel bottle well.
One of the T-piece ports must be vented to atmospheric pressure.
2Select ‘MEASURE’ from work mode item in “AG SETUP” menu, then set pump rate ‘low’
and wait for 10 minutes after the warm up information disappears.
3Enter ‘CALIBRATE’ menu, then open AG bottle and press the ‘VERIFY ACCURACY’ item.
Figure 5-9 AG Check Menu
4Observe the display value after 1 minute. The agent concentration accurate should be less than
±5%.
5
Choose other pump rate ‘middle’ or ‘high’and repeat the previous procedures.
(pump rate definition: three pump rate under adult mode: 100/150/200ml/min; neonate:
70/90/110 ml/min
6If the accurate over range, please press ‘START CAL’.

Table of Contents

Related product manuals