Chapter 3 Product Specifications
45
5 to 10
±0.4
10 to 15
±0.6
15 to 18
±1
>18 Not specified
0 to 1
±0.15
1 to 5
±0.2
5 to 8
±0.4
Sev
> 8 Not specified
0 to 1
±0.15
1 to 5
±0.2
Enf, Iso, Hal
> 5 Not specified
0 to 25
±1
25 to 80
±2
O2 (Optional)
80 to 100
±3
2 to 60 BrPM
±1 BrPM
AwRR
> 60 BrPM Not specified
Alarm range
CO2:
AwRR:
0 to 10 % (0 - 76 mmHg)
2 to 100 BrPM
Apnea alarm delay AwRR 20 - 40 s
Refreshing rate 1s
Calibration Yearly calibration requested.
Calibration stability
After module being used for 12 consective months, the error is <
1%
CO
2
250 ms (fall time 200 ms)
N
2
O 250 ms
O
2
600ms
HAL, ISO, SEV, DES 300 ms
Rise time (10 % to 90 %)
Sampling flow 120ml/min,
using the DRYLINE™
water trap and neonatal
DRYLINE™ sampling
line (2.5m)
ENF 350 ms
CO
2
250 ms (fall time 200 ms)
N
2
O 250 ms
O
2
500ms
Rise time (10 % to 90 %)
Sampling flow 200ml/min,
using the DRYLINE™
water trap and adult
DRYLINE™ sampling
HAL, ISO, SEV, DES 300 ms