98
Service Manual Combics
Corner load filing for single-cell Combics 60.000d platforms and scales
(CAAPP1-...-S, CAW.P1-...-S)
In order to reduce weighing platform deviations for off-center loads (cornerloads),
strain-gage load cell can be filed down.
Identifying corner load errors is routine for trained service technicians and is essen-
tial in ensuring that corner load correction is professionally implemented.
Generally speaking, load cell corner loads can only be improved if the deviations
on opposite corners have different signs (i.e. positive and negative).
Adjustment data