30
Service Manual Combics
Test Program for Combics Indicators| 1 and | 2
A test program is included on the main/digital PCB of the Combics indicator. When this
test is run, the action being carried out or the action to be carried out by the user is shown
on the display. A line with more than 7 characters is shown via scrolling characters on the
display.
The Combics 1 test program runs in the following sequence:
Information indicating how individual tests can be canceled
Version number information
Test of the display
Test of the keypad
Test of internal components (e.g. memory components)
Test of options and L9 (battery)
Test of COM 1 interface
Test of optional UNICOM interface, if used
The Combics 2 test program runs in the following sequence:
Information indicating how individual tests can be canceled
Version number information
Test of the display
Test of LEDs
Test of the keypad
Test of internal components (e.g. memory components)
Test of options and L9 (battery)
Test of COM 1 interface
Test of optional UNICOM interface, if used
Test of internal clock (RTC)
Operating Design | Test Program Combics 1 | 2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec Combics Series and is the answer not in the manual?

Minebea Intec Combics Series Specifications

General IconGeneral
BrandMinebea Intec
ModelCombics Series
CategoryScales
LanguageEnglish

Related product manuals