7.13.3 Parameters for the xBPI-weighing function
The following parameters must be entered for this menu item:
- Timeout for tare function depending on the application
- SBN-address for each xBPI-scale in bus operation mode
[Tare timeout]
Timeout for a zeroset or tare command to be executed.
If the xBPI scale has not executed the command in the specied time, the action will
be aborted.
Setting range: 09.9 s
[Serial number]
Serial number of the connected xBPI scale/weighing module.
With serial number "0", checking is omitted.
Setting range: 099999999
[SBN Address]
When the address is not set to 0, bus operation is active. Possible addresses: 1–31,
i.e., max. 31 xBPI scales can be operated on an RS-485 branch.
1. Select -[Weighing point]- [xBPI-Scale] and conrm.
2. Press the [Cong] softkey.
3. Enter the following parameters.
Transmitter Series PR 5220 7 Commissioning
Minebea Intec EN-98