7 Commissioning
7.1 Power failure/Data backup/Restart
7.1.1 Power failure
In the event of a power failure,
- all entered conguration and calibration data and
- all materials saved in the hard drive memory
are retained.
7.1.2 Data backup
The calibration data and parameters as well as all conguration and interface data are
stored in non-volatile (EAROM) memory.
Unauthorized data changing can be prevented by an access code.
Additional write protection is provided for calibration data and parameters (see
Chapter7.1.3.1).
7.1.3 Overwrite protection
7.1.3.1 CAL switch
The CAL switch protects the calibration data and parameters against unauthorized access.
The CAL switch is located under a cover that can be opened by means of a knife, see
Chapter 7.1.3.2.
When the CAL switch is in "opened" position, the calibration data and parameters can be
changed using the PC program or via the ProBus connection.
With the CAL switch in the "closed" position, the calibration data (e.g. dead load, span)
and parameters (measuring time, zero tracking etc.) cannot be changed.
7 Commissioning Transmitter Series PR 5220
EN-57 Minebea Intec

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec PR 5220/06 and is the answer not in the manual?

Minebea Intec PR 5220/06 Specifications

General IconGeneral
BrandMinebea Intec
ModelPR 5220/06
CategoryTransmitter
LanguageEnglish

Related product manuals