The position of the CAL switch is shown with VNC/WEB browser under [Show
status]:
[opened] = opened, no write protection.
[closed] = closed, write protection is active.
7.1.3.2 Removing the cover
Remove the cover as follows:
7.1.3.3 Factory settings (defaults)
Calibration data <default> Calibration parameter <default>
Upper scale limit (Max) <3000>
<Kg>
Measuring time (M) <320> ms
Scale interval <1> Measure rate <160> ms
Dead load <0.000000> mV/V Standstill time <1> M
Span <1.000000> mV/V Standstill range <1.00> d
Standstill timeout (timeout) <8> M
Calibration parameter <default> Test mode <absolute>
Overload (range > Max) <9>d Zero-setting range <50.00> d
1. Insert a knife (1) into the slot below the cover and press until it is released.
2. Remove the cap.
Transmitter Series PR 5220 7 Commissioning
Minebea Intec EN-58