Note:
Error messages, calibration see Chapter 16.5.
7.12.6 Determining the dead load
Note:
If a linearization was carried out (see Chapter 7.12.11), the following note appears on the
display after the [Dead load at] line is selected:
Cannot be changed here
while linearization is active.
Changes cannot be made while linearization is switched on.
Only deleting of the linearization points deactivates the linearization mode!
To use the empty scale/hopper as dead load (normal case):
1. Select [Scale interval] and conrm by pressing or .
A selection window opens.
2. Select the scale interval (1 d) and conrm with .
The verication is displayed by "Setting Scale interval".
The scale interval (d) is then calculated, based on the Max weight value.
1. Clear the scale/hopper.
2. Press the [by load] softkey.
3. Conrm entries with or .
The verication is displayed by "Setting dead load…".