7.12.12 Calculating the test value
The calculation of the test value is called by pressing the [CalcTest] softkey.
The maximum load (Max) is displayed with the ID without a weight unit. The value
determined during calibration by pressing the [CalcTest] softkey after starting the test is
displayed.
Depending on the settings under -[Weighing point]- [Calib]- [Param]- [Test mode]
the following is displayed by calling the Test with the key later on:
- with [Absolute] the maximum load
- with [Relative] the deviation from the test value
7.12.13 Saving the calibration
Quit calibration by pressing the softkey.
A message is displayed, indicating that the value for Max cannot be changed as
long as linearization is active.
7 Commissioning Transmitter Series PR 5220
EN-89 Minebea Intec