Note:
The number of vessel feet and the number of load cells may dier,e.g.: 4 vessel feet
on 1 pivots and 3 load cells.
[Max]
The load cell capacity is suggested as Max (E
max
× number of load cells).
The maximum load (Max) determines the maximum measured weight without dead
load. Normally, the selected Max must be smaller than the load cell capacity
(maximum capacity × number of load cells) – dead load, in order to prevent
overloading the load cells.
Enter the maximum load with decimal places (in this example: 200.000 t).
The key can be pressed to switch between units.
[Scale interval]
Selecting the scale interval (1 d) (in this example: 0020).
The scale interval (d) is calculated, based on the maximum weight value.
[Dead load]
To use the empty scale as dead load (normal case):
- Do not load scale.
- Press the [by load] softkey.
Note:
If the dead load is known, the value can be overwritten [by value].
[CAL weight]
- Center the CAL weight on the scale and enter the weight value with decimal
places (here: 11.000 t).
- Press the [Ok] softkey and remove CAL weight.
[Corner correction]
Perform a corner correction if necessary; see Chapter 7.14.10.3.
Note:
During calibration the weight can be displayed with 10 fold resolution by pressing
the key.
After 5 seconds the display returns to normal resolution. Press if you want to
switch to normal resolution immediately.
5. Press the softkey to exit the menu.
7 Commissioning Transmitter Series PR 5220
EN-117 Minebea Intec