EasyManua.ls Logo

Minebea Intec PR 5220/06 - 8.3 Configuration via a notebook;PC

Default Icon
235 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C
Max. load including dead load (load cell, cannot be changed)
The colors have the following meanings:
Red
Weight value is above maximum load (overload) or below -¼d
Green
Weight value is within tolerances
Orange
Weight value is above maximum capacity E
max
(max. capacity of load cell)
[No ID]
The Serial number is hidden.
[Show ID]
The Serial number is visible.
2. Select the desired load cell and press the [Info] softkey.
Display Description
Model name e.g.: PR6204/50tC3
Software version Software version of the load cell
Load cell serial number Serial number of the load cell
E
max
Maximum capacity
n Max. resolution
Y Minimum LC verication
Z Minimum preload signal recurrence
Overload Weight value above max. load
Overload counter Number of weight values above max. load
The higher the number, the higher the probabi-
lity of a faulty load cell.
The load cell data is displayed:
7 Commissioning Transmitter Series PR 5220
EN-155 Minebea Intec

Table of Contents