Instructions for Use SonaStar
®
Ultrasonic Surgical Aspiration System
IFU-601 Rev T March 2023 Page 55 of 73
4. During electrosurgery use, the needle may develop eschar buildup. To maximize efficiency, the tip
should be wiped clean with a sterile pad and alcohol. Briefly activating vibration with the tip
immersed in sterile physiological saline will also clean the tip and flush the vacuum system lines.
needle with abrasive or metallic
in device failure.
5. After use, the system may be dismantled and the Single-Use Monopolar Handpiece Cable should be
disposed of as per Hospital’s Biohazard Protocol. The umbilical cable is reusable and should not be
discarded.
6. Follow the electrosurgical generator manufacturer’s guidelines for dismantling and storing the unit.
7. Proceed to the next section to prepare the footswitch and associated components for storage.