Instructions for Use SonaStar Ultrasonic Surgical Aspiration System
IFU-601 Rev T March 2023 Page 7 of 73
Note
Inspect the output end of the handpiece where the probe is attached. Make sure this
surface is clean and has no scratches or marks. Inspect the threads for signs of damage
or excessive wear. Do not use if it appears damaged.
Note
The electrosurgery umbilical cable should be inspected for cuts, cracks, connector
damage, or any other signs of damage before use. Do not use the cable if it appears
damaged.
Note The SonaStar system should be fully tested and inspected prior to each procedure. The
console, footswitch, handpieces, all cables and accessories should be examined for
proper appearance and condition, and the initial system setup test should be
performed with each handpiece to ensure proper operation.
Note
The pinch valve should be checked for proper operation during setup. Proper tubing
routing and pinch valve operation should be checked.
Note
The
SonaStar
is
configured for electrical input (line) at the factory before shi
pment,
and is not intended to be configured or changed in the field except by Misonix, LLC
authorized technical personnel. The unit may only be used with the electrical input
originally intended.
Note
A second torque wrench is highly recommended to al
low continued system usage
during wrench recalibration. See Chapter 8 for more information on torque wrench
recalibration.
Note If the footswitch’s activation LED does not illuminate, plug the Remote IR receiver into
its receptacle at the console rear and place it within direct line of sight with the
footswitch. The Remote IR receiver can be moved closer to the wireless footswitch or
linked directly to it to improve signal reception as needed.
Note
If COAG is to be performed, the Single
-
Use Monopolar H
andpiece Cable must be
introduced into the sterile field, because the cable is supplied sterile, and is NOT
AUTOCLAVABLE. DO NOT RE-STERILIZE THE SINGLE-USE MONOPOLAR
HANDPIECE CABLE.
Note The SonaStar system has been designed for use with monopolar COAG using a
Misonix-approved electrosurgical generator. CUT MODE HAS NOT BEEN APPROVED
FOR USE WITH THE SonaStar. DO NOT USE CUT MODE WITH THIS DEVICE.
Note
Do not clean the tip with abrasive or metallic materials, as damage to the tip may result
in device failure.