16
EN
B CLASSIC / B FUTURA / S CLASSIC - TECHNICAL MANUAL
ALARMS
2. ALARMS
2.1 ERROR ALARMS - CODE “E”
CODE ALARM DESCRIPTION INDICATION ON LCD MODE
ERRORS (category A)
E 000 Blackout BLACKOUT 2
E 001 Voltage of power supply line too high OVERVOLTAGE 1
E 002 Water conductivity threshold 1 exceeded
H2O QUALITY INSUFFICIENT -
CHANGE CARTRIDGE PURE (*)
1
E 003 Water conductivity threshold 2 exceeded QUALITY H2O BAD 1
E 004 Network power frequency reading error NETWORK FREQUENCY ERROR 1
E 010 Door open DOOR OPEN! 1
E 020
Door lock system (closing) activation
timeout exceeded
DOOR LOCK TIMEOUT
1 (then reattempt
or turn off)
E 021
Door lock system (opening) activation
timeout exceeded
DOOR LOCK TIMEOUT
1 (then reattempt
or turn off)
E 022 Door lock microswitches failure DOOR LOCK PROBLEM 2
E 030
The water in the feed tank is at minimum
level
MIN WATER LEVEL 1
E 031
Maximum level of water in the drainage
tank
MAX DRAIN LEVEL 1
E 042
The MAX water level in the lling tank has
been reached
MAX FILLING LEVEL 1
E 900 Vacuum test failed (during TEST PHASE) TEST FAILED 2
E 901
Vacuum test failed (during STAND-BY
PHASE)
TEST FAILED 2
E 902
Vacuum test failed (vacuum pulse timeout
exceeded)
TEST FAILED 2
E 998 Access from remote in Service Mode REMOTE SERVICE 1
E 999 Manually interrupting the cycle MANUAL INTERRUPTION 2
2.2 FAULT ALARMS - CODE “A”
List of alarm codes, messages displayed on the LCD and relative RESET modes:
CODE ALARM DESCRIPTION INDICATION ON LCD MODE
ALARMS (category A)
A 032 Tank level sensors inconsistency H2O LEVEL PROBLEM 1
A 040 Automatic tank lling failed TOP-UP PROBLEM 1
A 042
WARNING: After feed tank max level has been reached with
Automatic Filling or Pure, stop the device
TOP-UP PROBLEM 1
A 101
Temperature sensor PT1 failed
(sterilization chamber)
PT1 INTERRUPTED 1
A 102
Temperature sensor PT2 failed
(steam generator)
PT2 INTERRUPTED 1
A 103
Temperature sensor PT3 failed
(tube bundle heating element)
PT3 INTERRUPTED 1
A 105
Temperature sensor PT5 failed
(conductivity measurement compensation)
PT5 INTERRUPTED 1
A 111
Temperature sensor PT1 short-circuited
(sterilization chamber)
PT1 SHORT-CIRCUITED 1