TECHNICAL MANUAL - B CLASSIC / B FUTURA / S CLASSIC
FUNCTIONING
1. FUNCTIONING
1.1 STANDBY
Upon power-on, the autoclave goes into STANDBY mode. User-selectable standby options are as follows:
- STANDBY OFF
- STANDBY ON
In the STANDBY OFF mode, the following components are powered off:
- solenoid valves;
- fans;
- water pump;
- vacuum pump.
The following components are not turned on:
- chamber band heater;
- generator heating element.
In the STAND BY mode the chamber heater is powered on and controlled by a heat controller that maintains:
- chamber temperature at 50°C with the door open
- chamber temperature at 100°C with the door closed
On STAND BY ON mode and door closed solenoid valve EV5 is activated.
STAND BY - components on state
STATE
EV1
Drain
High (NO)
EV2
Air inlet
(NC)
EV3
Drain Low
(NC)
EV4
Drain
bypass
(3V)
EV5
Air inlet to
chamber
(NC)
FAN
WATER
PUMP
VACUUM
PUMP
CHAMBER
BUNDLE HTR
GENERATOR
HTNG ELMT
OFF
Off
open
Off
closed
Off
closed
Off (0)
Off
closed
OFF OFF OFF OFF OFF
ON
door open
Off
open
Off
closed
Off
closed
Off (0)
Off
closed
OFF OFF OFF ON OFF
ON
door
closed
Off
open
Off
closed
Off
closed
Off (0)
On
open
OFF OFF OFF ON OFF