USER SAFETY INFORMATION
16
Cautions
Cleaning must be performed with the system turned off. Let all parts dry well before turning the power
back on.
Prevent liquids from penetrating the system, components, and transmitters. Do not spray the system with liquid
cleaning agents. If liquids have penetrated the system, open by authorized personnel for inspection and let dry
completely.
Do not attempt to clean the patient monitor or patient cables by submersing into a liquid, autoclaving, or steam
cleaning as this may damage equipment or reduce its usable life. Wipe the exterior surfaces with a warm water
and mild detergent solution and then dry with a clean cloth. Use of unspecified cleaning/disinfecting agents,
failure to follow recommended procedures, or contact with unspecified materials could result in increased risk of
harm to users, patients and bystanders, or damage to the patient monitor.
No user-serviceable parts inside. Screw removal by authorized service personnel only. Damaged or suspected
inoperative equipment must be immediately removed from use and must be checked/repaired by authorized
service personnel prior to continued use.
The rechargeable internal battery is a sealed lithium ion type. If the battery appears to become defective, refer
to Mortara Technical Support.
Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. When not in
use, patient cables can be stored. Keep patient cables should be stored off of the floor away from bedrails and
wheels to avoid cable damage. Roll the patient cables into a loose loop prior to hanging for storage.
When necessary, dispose of the patient monitor, its components and accessories (e.g., batteries, cables,
electrodes), and/or packing materials in accordance with local regulations.
Do not connect the patient monitor to any unauthorized patient monitors or use any third-party accessories. This
may cause inaccurate measurements or harm the patient. Installation and connection to data networks must be
performed by properly trained personnel, authorized by Mortara.
Check that all operating and environment conditions such as ambient temperature meet the specifications of the
Surveyor.
Do not exert excessive pressure on the touch panel LCD. Excessive pressure may permanently damage the
display.
During MRI scanning, the module must be placed outside the MRI suite. When the module is used outside the
MRI suite, EtCO2 monitoring can be implemented using the FilterLine XL.
Use of a CO2 sampling line with H in its name (indicating that it is for use in humidified environments) during
MRI scanning may cause interference. The use of non H sampling lines is advised.