Administration Setup Service Dialogue
The Service dialogue is a password protected area intended for use by service and system administrative personnel
only. The Setup Service dialogue is where the demonstration mode for simulated data on the patient monitor is
enabled and disabled. This dialogue is also where the monitor settings can be transferred to and from a USB thumb
stick when configuring multiple monitors. The monitor can be returned to the factory default configuration settings
in this dialogue. Restore the factory defaults to the unit by tapping the “Setup” icon, select “Administration”,
select
“service” enter the passwords, Dial 1 = 53, Dial 2 = 51, Dial 3 =54. Press “OK”, tap the ‘Restore Factory
Defaults”
icon and press “Ok”. You will see an on screen message that says “Please restart monitor for changes to
take
effect”. Power cycle the unit, and start a monitoring session with a new patient. If required, refer to S12/S19
User
Manual, 9515-183-50-xxx for that process.
Administration Setup Factory Dialogue
The Factory dialogue is a password protected area intended for use by service and administrative personnel only.
Contact service or system administrator for further details.