26
CAUTION: Keep the patient accessories off of the floor. Accessories that fall on the floor should
be inspected for defects, contamination, proper functionality, and cleaned or discarded according to the
approved recommendations.
CAUTION: The user has the responsibility to validate any deviations from the recommended
method of cleaning and disinfection.
Approved Cleaning Agents
Clean the touch screen with a soft cloth moistened with either a solution of 70%
isopropyl alcohol in distilled water or soapy water.
Do not spray cleaner directly onto the touch screen. Spray the cleaner onto a lint-free cloth
and then wipe the monitor.
To clean the touch screen display,
1.
Select the Settings sidebar button.
2.
Select the Administrative menu.
3.
Select the Screen Cleaning mode. This action disables the monitor’s touch screen for
15 seconds for cleaning purposes. After the 15 seconds expires, the touch screen
controls are reactivated.
Approved Cleaning Agents
Enzymatic detergent such as ENZOL (US) or CIDEZYME (outside the US)
Distilled water
Disinfectant solution (such as CIDEX OPA, or a 10% solution of household bleach
(5.25% sodium hypochlorite) in distilled water)
Soft, lint-free cloths and/or soft-bristled brushes
Protective gloves and eyewear
Procedure
1.
Disconnect the patient monitor from the wall outlet.
2.
Put on gloves and protective eyewear.
3.
Prepare the enzymatic detergent according to the manufacturer's instructions, and also
the disinfectant solution, in separate containers.
4.
Apply detergent to product using a soft, lint-free cloth. If material is dried on, allow
to sit for 1 minute. Do not immerse cable ends or lead wires in liquid as it can cause
corrosion.
5.
Wipe smooth surfaces with the cloth.
6.
Use a soft-bristle brush on visibly soiled areas and irregular surfaces.
7.
Remove detergent from product using cloth dampened in distilled water.
8.
Repeat as necessary.
9.
Apply disinfectant solution on affected area using a soft cloth. Allow product to sit
for 5 minutes.
10.
Wipe excess solution and clean product again with cloth dampened in distilled water.
11.