49
Alarm Silence Time
This setting controls the length of silence time for the Alarm Silence quick action key on the main screen.
To select the Alarm Silence Time:
1.
Select the Setup quick action key on the main screen to access the Setup dialogue.
2.
Select the Administration button.
3.
Select the Alarms button to access the Administration: Setup Alarms dialogue.
4.
Locate the Alarm Silence Time label.
5.
Select the arrow to display the drop down list box (Choices are 1 minute, 1.5 minutes, 2 minutes,
5 minutes and 10 minutes).
6.
Select the alarm silence time desired.
7.
Select the OK button to enter that change OR select the Cancel button to cancel the changes.
Alarm Suspend Time
This setting controls the length of silence time for the Alarm Suspend function.
To select the Alarm Suspend Time:
1.
Select the Setup quick action key on the main screen to access the Setup dialogue.
2.
Select the Administration button.
3.
Select the Alarms button to access the Administration: Setup Alarms dialogue.
4.
Locate the Alarm Suspend Time label.
5.
Select the arrow to display the drop down list box (Choices are 1 minute, 1.5 minutes, 2 minutes,
5 minutes and 10 minutes).
6.
Select the alarm silence time desired.
7.
Select the OK button to enter that change OR select the Cancel button to cancel the changes.
Can Suspend All Alarms
This setting controls if the clinician has the ability to suspend all alarms including the lethal arrhythmias
(Asystole, VTach, and VFib).
To configure:
1.
Select the Setup quick action key on the main screen to access the Setup dialogue.
2.
Select the Administration button.
3.
Select the Alarms button to access the Administration: Setup Alarms dialogue.
4.
Locate the Can Suspend All Alarms label.
a.
Selecting Enabled means the clinician can suspend all alarms including the lethal arrhythmias.
b.
Selecting Disabled means the clinician cannot disable all alarms. If this choice is selected, the
Suspend All Alarms button will be hidden from the Setup dialogue.
5.
Select the OK button to enter that change OR select the Cancel button to cancel the changes.
Can Disable HR/PR Alarms
This setting controls if the clinician has the ability to disable the HR/PR alarms.
To configure:
1.
Select the Setup quick action key on the main screen to access the Setup dialogue.
2.
Select the Administration button.
3.
Select the Alarms button to access the Administration: Setup Alarms dialogue.
4.
Locate the Can Disable HR/PR Alarms label.
a.
Selecting Enabled means the clinician can disable the HR/PR alarms.
b.
Selecting Disabled means the clinician cannot disable the HR/PR alarms.
5.
Select the OK button to enter that change OR select the Cancel button to cancel the changes.