MSA Using the Compressed Air Breathing Apparatus
AirGo
35
3.8. Use of Additional Connections for Medium Pressure
(1) Remove protection cap from the coupling of the additional connection for
medium pressure.
(2) Connect medium pressure line of lung governed demand valve of second
user until the coupling audibly catches.
Warning!
When rescuing persons with the rescue set via the second connection,
more air is consumed.
Hence, the service time is considerably reduced. Always keep this in
mind when using your apparatus.
3.9. Handling the Warning Device
Fig. 22 Manifold
1 Warning signal with protective cap
2 Second connection
Only applicable to SingleLine Pneumatics.
After using, it is possible to reduce the volume of the warning device during the
decontamination process. This is done by removing the protective cap from the
second connection on the manifold and pushing it onto the warning device.
Warning!
During duty, damping down the warning signal tone is not permitted.
Remove the protective cap from the warning device again and push it
onto the second connection after the SCBA Basic Apparatus has been
removed.