32 ALTAIR 5 - Operating Manual
3.4.1 Calibration Options
Monochrome display Colour display
CALIBRATION OPTIONS CALIBRATION OPTIONS
The Calibration Options menu allows the user to:
modify the calibration cylinder settings
(CYLINDER SETUP)
enable/disable calibration due and to set the number of days
(
CAL DUE OPTIONS)
enable/disable the option to show the last cal date at turn on and
(LAST CAL DATE)
enable/disable the option for password protected calibration
(
CAL PASSWORD)
Press
the
button go to next page
the
button to go previous page
the
button to enter setup.
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
(1) Press the button to enter setup.
The screen for the first calibration cylinder displays.
(2) Press
the
or
button to change the value.
the
button to confirm the setup.
With this confirmation the instrument automatically moves to the next cylinder
setting.
(3) Repeat the sequence for changing the required settings for all necessary
gas values.
After the last setting is performed, the instrument returns to the
Calibration Options menu.