EasyManua.ls Logo

MSA altair 5 - IR Sensors

MSA altair 5
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GB
TECHNICAL SPECIFICATIONS/CERTIFICATIONS MSA
54 ALTAIR 5 - Operating Manual
5.3. IR sensors
Gases Display range Smallest
resolution
Response time
at 20°C
Reproducibility
of the zero point
Reprodu-
cibility
of the
measured
value
1)
t
50
t
90
CO
2
0 – 5 % Vol 0.01 % Vol
≤ 12 s ≤ 36 s ≤ ± 0.05 % Vol ≤ ±5%
CO
2
0 – 10 % Vol 0.01 % Vol
≤ 12 s ≤ 35 s ≤ ± 0.1 % Vol ≤ ±4%
CO
2
0 – 50 % Vol 0.5 % Vol
≤ 14 s ≤ 36 s ≤ ± 1.0 % Vol ≤ ±4%
C
3
H
8
0 – 100% LEL 1 % LEL
≤ 11 s ≤ 32 s ≤ ± 3 % LEL ≤ ±8%
C
3
H
8
0 – 25 % Vol 0.5 % Vol
≤ 12 s ≤ 35 s ≤ ± 0.5 % Vol ≤ ±4%
C
3
H
8
0 – 100 % Vol 1 % Vol
≤ 12 s ≤ 36 s ≤ ± 3 % Vol ≤ ±8%
C
4
H
10
0 – 100% LEL 1 % LEL
≤ 11 s ≤ 32 s ≤ ± 3 % LEL ≤ ±8%
C
4
H
10
0 – 25 % Vol 0.5 % Vol
≤ 12 s ≤ 35 s ≤ ± 0.5 % Vol ≤ ±4%
C
4
H
10
0 – 100 % Vol 1 % Vol
≤ 12 s ≤ 36 s ≤ ± 3 % Vol ≤ ±8%
CH
4
0 – 25 % Vol 0.5 % Vol
≤ 12 s ≤ 35 s ≤ ± 0.5 % Vol ≤ ±5%
CH
4
0 – 100 % Vol 1 % Vol
≤ 11 s ≤ 34 s ≤ ± 5 % Vol ≤ ±10%
1) Data in % of measurement range