GB
MSA OPERATION
ALTAIR 5 - Operating Manual 35
Setting Sensor Alarms
This page allows modifying the preset alarm values of:
• LOW Alarm
• HIGH Alarm
• STEL Alarm
• TWA Alarm.
(1) Press the
button to enter Sensor Alarm setup.
• LOW Alarm Setup screen displays.
COMB
CO
10
35
2.0
PPM
PPM
Vol%
10
PPM
SETUP
LOW ALARM
CHANGEABORT
NEXT
LEL
19.5
Figure 3-5 Sensor Alarm Setup
(2) Press
the
button to abort the operation or
the
button to go to next alarm setup or
the
button to change the alarm setpoints.
• Alarm Value for the first Sensor displays.
COMB
CO
10
35
2.0
PPM
PPM
Vol%
10
PPM
SETUP
OK
19.5
Figure 3-6 Sensor Alarm Setup
(3) Set values for Sensor Alarm by pressing the
or
button.
(4) Press the
button to confirm set values.
(5) Repeat setting for all other sensors.
(6) Press the
button to return to the Alarm Options menu.
(7) Repeat setting for all other alarm types.