GAS TYPE RANGE SPAN GAS MSA RP CALIBRA- WARM-UP
PRESET CYLINDER TION KIT TIME
VALUES P/N
Carbon Dioxide IR 0-5000 PPM 2000 PPM 479266 40 ---------
0-2% 1.5% 807386 40 ---------
0-5% 2.5% 479265 40 ---------
Hydrogen Chloride 0-50 PPM 40 PPM 10028078 54 30 minutes
NOTES:
1
Calibrated with PROPANE (.6% GAS BY VOLUME)
2
Not required for standard calibration procedure
3
For Combustible Gas, it is good practice to calibrate unit with gas to be detected
4
ClO
2
is calibrated with Cl
2
or use ClO
2
Calibrator Kit (P/N 710420)
5
Methane IR is calibrated with 50% LEL Methane
6
ETO is calibrated with SO
2
.
7
Hydrogen Fluoride (HF) is calibrated with Sulfur Dioxide (SO
2
), 10 PPM SO
2
= 8 PPM HF
Figure 2-1. Calibration Kit 40 Contents
(Your Kit may also include one or two gas cylinders)
Item 2 - Zero Cap
(P/N 710535)
Item 3 - 1.5 LPM Flow
Controller (P/N 478358)
Item 1 - Tubing (P/N 711112)
- 3/16" ID side connects to Item 3
- 1/4" ID side connects to sensor
2-4

Table of Contents

Other manuals for MSA Ultima Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA Ultima Series and is the answer not in the manual?

MSA Ultima Series Specifications

General IconGeneral
DisplayLCD with backlight
Communication PortsRS-485, Modbus RTU
Measurement RangeDepends on sensor connected
AccuracyDepends on sensor connected
Outputsrelays
CertificationsUL

Related product manuals