34
If your control solution test result is out of the range, repeat the
test. If the results from the control solution tests continue to fall
outside that range, do not use the MyStar Extra® Meter to test your
blood glucose. Call Customer Service at 08000 35 25 25. The phone
number is also printed on your meter label.
IMPORTANT: Out-of-range test results may be caused by one or
more of the following factors:
[1] Expired or defective control solution.
[2] Expired or defective test strip.
[3] Error in performing test.
[4] Watered-down control solution.
[5] Meter malfunction.
[6] Failure to shake the control solution bottle vigorously before
using.
[7] Failure to discard the  rst drop of control solution and to wipe
the bottle tip clean.
[8] Control solution test done outside the system operating
temperature range of 1C to 40°C.
Performing a Control Solution Test