Performance Verification
30
2. Select 60 BPM on the SRC-MAX.
3. Connect the 3-Lead ECG cable to the ECG monitor and
the ECG port of the SRC-MAX.
4. Select Lead 1 on the ECG monitor.
5. Verify the 60 BPM reading on the ECG monitor.
6. Repeat step 5 for Lead selection 2 and 3.
7. Select 200 BPM on the SRC-MAX.
8. Select Lead 1 on the ECG monitor.
9. Verify the 200 BPM reading on the ECG monitor.
10. Repeat step 9 for Lead selection 2 and 3.