Performance Verification
SRC-MAX 29
Figure 17: 200 BPM
9. The period of the signal should be 0.27 to 0.33 seconds.
10. The pulse width of the signal should be 50 to 100
milliseconds.
BPM Test (Alternate Method)
Equipment Required:
• Pulse oximeter or monitor
• ECG monitor
• Three lead ECG cable
Setup
Tolerances for this procedure are in the Specifications section,
page 37.
1. Connect the SRC-MAX to the powered-up pulse
oximeter or monitor. The SRC-MAX should power up
normally.