Environment Description
Operating Humidity 15 to 95% non-condensing
Operating Altitude Sea level to 15,000 ft (0 to 4,572 m)
There is no altitude limitation when
HT70 is operated in a pressurized
environment.
Operating Pressure 600 to 1,100 mbar
Storage and Shipping Temperature
–40ºC to 65ºC
Storage and Shipping Humidity 0 to 95% non-condensing
Size and Weight
Height (includes handle) 10.25 inches (26.04 cm)
Width 9.75 inches (24.77 cm)
Depth 11 inches (27.94 cm)
Weight 17.40 lbs. (7.89kg)
Factory Default Parameters
Patient Settings:
MODE A/CMV
VT (Volume Control) 500 mL
i time 1.0 sec
RR 15 b/min
Ptrig –1.0 cmH2O
iPaw Alarm 5 cmH2O
hPaw Alarm 40 cmH2O
hMVI Alarm 3 L/min
iMVI Alarm 20 L/min
PEEP/CPAP 0 cmH2O
PS 0 cmH2O
Alarm Loudness Level 7
Miscellaneous
Patient Circuit Reusable or disposable 22 mm I.D.
adult or 15 mm I.D. pediatric circuit with
3/16 inch (4.8 mm) I.D. proximal
pressure sensing line, 1/8 inch (3.2 mm)
I.D. exhalation valve control drive tubing,
and exhalation valve.
NOTE: Newport Medical cannot
guarantee the safe use of breathing
circuits that are not recommended by
Newport.
OPRHT70 A0310
9-8
9
Specifications