EasyManua.ls Logo

Newport HT70 - 8 Cleaning and Maintenance; Cleaning and Disinfecting; Ventilator

Newport HT70
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPRHT70 A0310
8-1
Cleaning and Maintenance
8
Cleaning and Disinfecting
Use the information in this section in conjunction with hospital policy,
physician prescription, Homecare Dealer or accessory manufacturer
instructions.
Definition of CLEAN: Employ a medical detergent or alcohol based
cleaning solution to remove blood, tissue and other residue. Rinse
thoroughly with sterile, distilled water, and allow to air dry.
Definition of DISINFECT: Employ liquid chemicals to disinfect.
Definition of Sterilize: Employ pasteurization, steam autoclave or
Ethylene Oxide (EtO) to sterilize.
Caution: When using liquid chemical agents, closely follow the
manufacturer’s recommendations. Prior to use, verify that the agent
is compatible with plastics.
WARNING Ethylene Oxide (EtO) is toxic. All accessories MUST
be completely dry prior to packaging for ethylene oxide sterilizing.
After sterilizing, they must be properly aerated to dissipate residual
gas absorbed by the material. Follow the EtO manufacturer’s
recommendations for the specific aeration periods required.
Caution: Ethylene Oxide (EtO) may cause superficial crazing
of plastic components and will accelerate the aging of rubber
components.
Caution: Always inspect breathing circuits and accessories after
cleaning, disinfecting or sterilizing to check for deterioration. If the
any part is damaged or shows excessive wear, replace with a new
part. Do not use cracked or damaged parts.
Ventilator
Wipe clean between patients and as needed while in use. The
exterior of the ventilator should be wiped clean with a cloth
dampened with a medical detergent, disinfectant or alcohol based
cleaning solution.
Caution: Do not use agents that contain acetone, toluene, halogenated
hydrocarbons, or strong alkalines on the face panel or ventilator housing.
Caution: Never autoclave or EtO sterilize the HT70 ventilator. These
processes will damage the HT70, rendering it unusable.

Table of Contents