OPRHT70 A0310
4-16
4
Navigating the Screens
2. Make all the changes on the ventilator that you require for a
selected patient group (infant, pediatric or adult). Be sure to
check the settings in the Advance screen and the Alarms
screen.
3. Once you are satisfied with the settings for your patient group,
return to the Advance screen ¢
Utility screen and touch
press the Set Custom Presets button.
4. Touch the desired Preset (infant, pediatric or adult).
5. Touch the Accept button to confirm and save your choice.
Prior to starting ventilation, you can choose “Activate Presets” button to
enable your preset parameters from the Startup Screen (see page 4-3).
NOTE: Custom Presets are retained after power down.
Trends Screen Navigation*
(Hospital Domain)
The Trends Screen displays trended data for monitored parameters.
Pushing the s or t panel buttons will move the cursor to the right or
to the left.
As the cursor line moves across the graph, the time display will
indicate the time at that point on the graph and the numbers on the
line will show the value for each monitored parameter.
The yellow vertical bars represent power On/Off cycling events.
There are three sets of parameters to display.
To Select a Trend Set:
To change the displayed parameters, touch the Trends Set button on
the right side of the touchscreen to scroll through these choices:
1) Peak pressure, Mean pressure and PEEP
2) Peak Flow, Power Pac, Backup Battery
3) BAT PAC Temp, BU BAT Temp, Internal Temp
4) Tidal Volume, Resp. Rate Total, Minute Volume
* Not available on all models