OPRHT70 A0310
9-2
9
Specifications
* Not available on all models
SIMV
Controls / Monitors Range / Selection Resolution
MODE (Pressure or A/CMV
Volume Control)
SPONT
NIV (Non Invasive
On or Off, When On, allows iMin Vol
ventilation)
Alarm to be set to Off, iP Alarm to be
to be set 1 cmH2O/mbar above PEEP
VT (Tidal Volume) 50 to 2,200 mL, ATPS, ± 10% 1.0 mL
PC
(Pressure Control)
5 to 60 cmH2O / mbar
1.0 cmH2O/mbar
VC (Volume Control) 0.05 - 2.20 L 0.01 L
Flow 6 to 100 L/min 1.0 L/min
i time (Inspiratory Time)
0.1 to 3.0 sec 0.1 sec
RR
(Respiratory Rate)
1 to 99 b/min 1.0 b/min
P trig (Sensitivity)
9.9 to 0 cmH2O / mbar, pressure triggering
0.1 cmH2O/mbar
PEEP/CPAP 0 to 30 cmH2O / mbar
1.0 cmH2O/mbar
PS (Pressure Support)*
0 to 60 cmH2O / mbar above baseline pressure,
1.0 cmH2O/mbar
limited to PEEP + Psupport 60 cmH2O / mbar
Controls/Monitors
Manufacturer Name and Address
O
i
h
s / t
Authorized Representative in the European community
Main Power Off / On (momentary switch)
Low (Paw or Min Vol) alarm
Equipotentiality
Applied Parts Type BF
Brightness control
Alarm Silence
Up / Down arrow
Federal law (US) restricts sale by or on the order of a physician
Meets FAA requirements in RTCA standard,
DO160, sec 21 category M for use in all stages of
air travel, including takeoff and landing.
High (Paw, Min Vol or RR) alarm
Attention, see instructions for use
EC REP
Miscellaneous Reference Symbols

Table of Contents