Flow Wave Pattern Square or Descending
OPR360U A0509
5-1
5
Controls / Monitors Range / Selection Resolution
I:E Ratio 1:599 to 3:1 0.1 for 9.9:1
to 1:99 and
1 for 99:1 to
1:99
Airway Pressure 10 to 100 cmH
2O / –10 to 98 mbar
Gauge includes indicator bars to show low
and high PAW alarm limits
Optional O
2 Sensor Enabled or Disabled
When Enabled, high and low FiO2
alarms are active
PS Max i time * 0.1 - 3.0 sec 0.1 sec
PS % Exp Threshold* 5 - 55% 5
Slope Rise 1 -10 (1 is slowest) 1
AutoLock function Enabled/ Disabled
AutoLock icon Touch for 3 seconds to unlock
buttons if AutoLock is Enabled in
Utility screen.
When Lock is on screen, all controls
are locked except Alarm Silence
Reset and Brightness control.
Monitor Data Selections
Front Panel Membrane Buttons and Indicators
Minute Volume 0.01 L to 50.0 L 0.01 L
Insp. Tidal Volume 0.01 L to 3.0L 0.01 L
RR total 0 to 200 b/min 1 b/min
P Peak 0 to 100 cmH
2O /mbar
1 cmH2O/mbar
P Mean 0 to 100 cmH2O /mbar
1 cmH2O/mbar
P Base (PEEP) 0 to 100 cmH2O /mbar
1 cmH2O/mbar
(Peak) Flow 5 to 150 L/min 0.1 L/min
FiO2 (optional) 0.21 to 1.00 0.01
I:E Ratio 1:99.0 to 3.0:1
0.1 (9.9:1 to 1:9.9)
1.0 (99:1 to 1:99)
Cancel Press front panel button to cancel any touchscreen
settings changes that have not been accepted
Accept Press to confirm or accept any touchscreen
setting changes
* Not available on all models
OPR360U A0509
5-1
5
OPRHT70 A0310
9-3
Specifications