Ventilator Alarms
6
See Section 9, Specifications, for alarm priority level, ranges and
descriptions.
Setting Alarms
The Alarm controls are changed just like the parameter controls -
with a simple Touch / Adjust / Accept method:
1. Touch the Alarms button to enter the Alarms screen.
2. Touch the desired Alarm control (it will appear highlighted)
3. Use the s or t buttons to make the desired adjustment.
4. Press the Accept button to confirm the change.
You can make several adjustments without accepting the changes.
When you are satisfied with all the changes, you can accept them all
by pressing the Accept button once.
Before you accept any change, if you want to go back to the previous
settings, simply press the Cancel button.
Alarm Quickset
You can choose to have the HT70 automatically set the alarm limits.
From the Alarms screen, touch Alarm Quickset and confirm by
pressing the Accept button. The HT70 will monitor ventilation for 30
seconds and then set the alarm limits. During the 30 second period
the touch screen will not respond unless an alarm occurs or the
Cancel button is pressed.
If an alarm occurs during the monitoring period, Quickset is
canceled. If this occurs, check the ventilator settings and confirm that
they meet the physicians prescription and are meeting the patient’s
needs. Then you can activate Quickset again.
Alarm Quickset will not activate when in standby condition, you must
be in ventilation condition.