Basic Domain
This is a simplified screen for use in a homecare or sub-acute facility
type environment. The central parameter platform is enlarged for
easier access. Setting changes are limited to the main parameters
and the alarm settings. The Breath Type, Mode and NIV settings are
displayed but cannot be set within this Domain. The displayed Monitor
Data buttons remain as they are set from the Hospital Domain and
cannot be changed. This helps simplify operation for users who don’t
need to change advanced settings or access technical information. If
access is needed for these settings, simply use the Domain button in
the lower right corner of the screen to toggle to the Hospital Domain.
Press Accept to confirm your selection.
Basic Domain Screen selections
Transport Domain
Transport Domain also uses an enlarged central parameter platform
for easier access and viewing. This Domain has full access to change
the Main parameter settings and the Alarm settings. In the Transport
Domain the user has the ability to change Breath Type, Mode, and
Main