OPRHT70 A0310
4-5
Navigating the Screens
4
Basic Domain
This is a simplified screen for use in a homecare or sub-acute facility 
type environment. The central parameter platform is enlarged for 
easier access. Setting changes are limited to the main parameters 
and the alarm settings. The Breath Type, Mode and NIV settings are 
displayed but cannot be set within this Domain. The displayed Monitor 
Data buttons remain as they are set from the Hospital Domain and 
cannot be changed.  This helps simplify operation for users who don’t 
need to change advanced settings or access technical information.  If 
access is needed for these settings, simply use the Domain button in 
the lower right corner of the screen to toggle to the Hospital Domain. 
Press Accept to confirm your selection.
Basic Domain Screen selections
Transport Domain
Transport Domain also uses an enlarged central parameter platform 
for easier access and viewing. This Domain has full access to change 
the Main parameter settings and the Alarm settings.  In the Transport 
Domain the user has the ability to change Breath Type, Mode, and 
Main
Screen
Alarm
Screen