Navigating the Screens
4
If the test fails:
o The message “Circuit Check FAILED Press Accept to Continue
will be displayed.
o Press the Accept button to return to Startup Screen.
o Check that the accessories and all tubing connections are
properly connected and leak free.
o Then touch the Circuit Check button to redo the test.
If the Circuit Check fails repeatedly, try a different circuit.
WARNING Do not use the HT70 if the Circuit Check fails, inadequate
ventilation may result. Use an alternate method of ventilation.
Contact Newport Medical Technical Support.
NOTE: The Circuit Check results are logged into the Event History
and retained after power down.
Activate a Patient Preset
Touch the Activate Preset button on the Startup Screen. A new
screen appears with the Patient Preset options. Touch the desired
Patient Preset and confirm by pressing the Accept button. Touching
the ‘Reset Event History’ button causes a “New Patient” record to be
added to the Event History Log.
See the Utility Screen Navigation section for instructions on how to
customize the Custom Patient Presets.
NOTE: If you do not want to use these Preset settings, press the
Cancel button to return to the Startup screen.