Explanation of Modes and Controls
10
A/CMV Mode
(Assist/Control Mandatory Ventilation)
In A/CMV mode, time activated (mandatory) breaths are delivered 
in accordance with the RR setting. Patients can trigger mandatory 
breaths in addition to, or in place of, time activated (mandatory) 
breaths if the effort they generate causes airway pressure to meet 
the Ptrig setting. Every such patient effort results in a mandatory 
breath. The breath can be volume or pressure controlled. PEEP/CPAP 
may be added. In Pressure Control, tidal volume is determined by 
the target pressure, i time, and patient respiratory mechanics and in 
Volume Control, by the tidal volume setting.
SIMV Mode
(Synchronized Intermittent Mandatory Ventilation)
In SIMV mode, patients receive a fixed number of volume or pressure 
controlled mandatory breaths (time or patient activated) and may 
breathe spontaneously between mandatory breaths, with or without 
pressure support (PS)*. PEEP/CPAP may be added.
The first patient triggered breath in any mandatory breath interval will 
be a patient triggered mandatory breath. The patient has the rest of 
the interval to breathe spontaneously. If the patient does not trigger 
the ventilator, and one complete mandatory breath interval has 
elapsed, a time triggered mandatory breath is delivered.
A mandatory breath lockout interval is activated whenever the patient 
triggers a mandatory breath. This limits the number of mandatory 
breaths (time triggered or patient triggered) the patient receives in 60 
seconds to the RR (b/min) setting.
SPONT Mode
(Spontaneous Ventilation)
In SPONT mode, mandatory breaths are not delivered but the user 
can adjust both PEEP/CPAP and pressure support (Psupport) levels. 
The patient has control over each breath.
When PEEP/CPAP is set above 0, the ventilator mode is CPAP 
(without PS) or Bi-level Positive Airway Pressure (with PS). Ensure 
that Ptrig is set so the HT70 detects all spontaneous patient efforts.