Loading...
OPR360U A0509
5-1
5
OPR360U A0509
5-1
5
OPRHT70 A0310
10-5
Explanation of Modes and Controls
10
Ventilation is canceled. If linked to Apnea Alarm, after 2 minutes
of Backup Ventilation it is canceled. At that time the audible alarm
stops, the alarm indicator latches and the HT70 resumes ventilation
at the user-selected parameters.
Press the Alarm Silence/Reset button to cancel the latched alarm
indicator and alarm message in the Message Display Window.