OPRHT70 A0310
4-11
Navigating the Screens
4
To enter Alarms Screen: Touch the Alarms button to open the alarm
setting screen. This same button can be used to toggle back to the
Main Screen in all Domains.
To set Alarms: Touch desired button to activate (highlight) an alarm
limit, then use the s or t panel buttons to adjust, then press Accept
to confirm changes or press Cancel to return to original settings.
1. Settable Alarms
#P (High Pressure) #MVI (High Minute Volume)
$P (Low Pressure) $MVI (Low Minute Volume)
#RR (High Respiratory Rate) Apnea (time adjustment)
#FiO2 (High FiO2) (optional)
$FiO2 (Low FiO2) (optional)
2. Alarm Loudness Level 1-10 (10 is loudest) – touch this button
then use the s or t panel buttons to adjust the loudness level then
press Accept. Always set the alarm loudness level high enough to
ensure that caregivers will hear any alarm. If necessary attach a
remote alarm system to ensure that the caregiver can always hear an
alarm when it sounds.
3. Alarm Quickset – Alarm Quickset will automatically set the alarm
limits. Touch this button to enter the Alarm Quickset screen, and then
follow the screen prompts. Press Accept to activate or Cancel to
return to the Alarms screen.
When selected, Alarm Quickset monitors settings for 30 seconds
and then sets the alarms. If an alarm occurs during the monitoring
period, Quickset is canceled. During the 30 second period the touch
screen will not respond unless an alarm occurs or the Cancel button
is pressed.
Alarm Quickset will not activate in standby condition.