EasyManua.ls Logo

Nikon ECLIPSE MA200 - Epi-fl Microscopy

Nikon ECLIPSE MA200
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Microscopy
22
Return to the bright-field microscopy under the episcopic illumination.
1 Pull the polarizer/analyzer unit to the first
stop-click position and remove the
polarizer/analyzer from the optical path. (See
3.13.1.)
2 Both the λplate and the polarizer/analyzer unit are
removed from the optical path. (See 3.14.1.)
3 Detach the DIC slider to remove the DIC prism
from the optical path. (See 3.15.1 or 3.16.2.)
4 Turn the brightness control dial to adjust the
brightness of the episcopic illumination. (See
3.2.1.)
5 Adjust the brightness with the desired ND filter in
the filter turret. (See 3.10.)
2.5 Epi-fl Microscopy
Attach the items required for the epi-fl microscopy to the microscope. (See Table 2.1.)
To perform the epi-fl microscopy using the C-HGFI/C-HGFIE HG Precentered Fiber Illuminator as the
external light source, screw the compensation filter supplied with the fiber adapter into the MA2-FL
fluorescent unit. (Refer to 4.15.1.)
Find the target and focus on the sample by BF/DF microscopy under the episcopic
illumination. (See Pages 16 to 17, and 19.)
Set the microscope for the epi-fl microscopy.
1 Turn the revolving nosepiece to place the
objective of a desired magnification into the optical
path. (See 3.9.2.)
2 Turn the BD field changeover lever to the “BF
(bright-field)” position. (See 3.3.)
3 Push the attached fluorescent unit into the second
click-stop position and place the fl filter into the
optical path (See 3.19.1.)
4 Turn the brightness control dial to adjust the
brightness. (See 3.2.1.)
5 Adjust the brightness with the desired ND filter in
the filter turret. (See 3.10.)
3-4
3-5
3-3
3-2
3-1

Table of Contents

Related product manuals